    Xin JIN, Luping WU, Tingting ZHU. Discussion on quality control of liver function reserve analyzer (ICG)[J]. Optical Instruments, 2024, 46(3): 40 Copy Citation Text show less
    Principle of liver function reserve analyzer
    Fig. 1. Principle of liver function reserve analyzer
    Light-emitting diode
    Fig. 2. Light-emitting diode
    Photodiode
    Fig. 3. Photodiode
    Experimental related equipment and reagents
    Fig. 4. Experimental related equipment and reagents
    Test results
    Fig. 5. Test results
    名称试验1试验2试验3试验4试验5试验6
    标准浓度100%80%60%40%20%10%
    探头1100%80%59.9%40%19.8%10.1%
    探头2100%79.7%59.3%39.5%19.6%9.4%
    探头3100%79.1%58.9%39.3%18.1%8.8%
    探头498.7%78.5%57.2%37.6%17.4%7.2%
    Table 1. Average detection test results
    Xin JIN, Luping WU, Tingting ZHU. Discussion on quality control of liver function reserve analyzer (ICG)[J]. Optical Instruments, 2024, 46(3): 40