• Journal of Applied Optics
  • Vol. 45, Issue 2, 276 (2024)
Xuzhou HE, Feng LIN*, Zhongxin XIE, Wei CHEN, Kangsheng WANG, Liyu LIN, and Junjie LIN
Author Affiliations
  • Fujian Provincial Key Laboratory of Photonics Technology, Key Laboratory of Optoelectronic Science and Technology for Medicine (Ministry of Education), College of Photonic and Electronic Engineering, Fujian Normal University, Fuzhou 350007, China
  • show less
    DOI: 10.5768/JAO202445.0201001 Cite this Article
    Xuzhou HE, Feng LIN, Zhongxin XIE, Wei CHEN, Kangsheng WANG, Liyu LIN, Junjie LIN. Design of endoscopic optical system with 2k resolution[J]. Journal of Applied Optics, 2024, 45(2): 276 Copy Citation Text show less
    Ideal optical model for simplified anti-telescopic objective structure
    Fig. 1. Ideal optical model for simplified anti-telescopic objective structure
    Initial structure diagram of objective lens
    Fig. 2. Initial structure diagram of objective lens
    Initial structure diagram of relay system
    Fig. 3. Initial structure diagram of relay system
    2D structure diagram of final designed objective module
    Fig. 4. 2D structure diagram of final designed objective module
    2D structure diagram of final designed relay system
    Fig. 5. 2D structure diagram of final designed relay system
    2D structure diagram of final designed endoscopic system
    Fig. 6. 2D structure diagram of final designed endoscopic system
    MTF curves
    Fig. 7. MTF curves
    Spot diagram
    Fig. 8. Spot diagram
    Distortion diagram
    Fig. 9. Distortion diagram
    参数指标
    F#<8
    工作波长/nm486~656
    全视场角/(°)75
    视向角/(°)≤30
    系统总长/mm<400
    物方分辨率/lp·mm−1≥30
    物方角分辨力/c/(°)>14
    像高/mm8.9
    Table 1. System parameters of endoscope
    蒙特卡洛样本百分比/%各视场平均值15.3°21.7°26.5°30.6°34.2°37.5°
    900.10470.12230.13010.12040.10840.09270.08440.0510
    800.13310.14400.16310.15690.14690.13530.11140.0862
    500.18470.20910.23290.22320.20660.18600.17230.1280
    200.21990.25880.26780.25830.24420.23440.22440.1648
    100.23810.28200.28220.27520.26050.24690.23080.2013
    Table 2. Results of Monte Carlo tolerance analysis
    公差数值
    曲率半径公差/rings2
    厚度公差/mm±0.03
    X方向表面偏心公差/mm±0.02
    Y方向表面偏心公差/mm±0.02
    X方向元件偏心公差/mm±0.02
    Y方向元件偏心公差/mm±0.02
    X方向表面倾斜公差/(°)±0.05
    Y方向表面倾斜公差/(°)±0.05
    X方向元件倾斜公差/(°)±0.05
    Y方向元件倾斜公差/(°)±0.05
    折射率公差0.001
    阿贝常数公差1
    Table 3. Allowable tolerances for parts
    Xuzhou HE, Feng LIN, Zhongxin XIE, Wei CHEN, Kangsheng WANG, Liyu LIN, Junjie LIN. Design of endoscopic optical system with 2k resolution[J]. Journal of Applied Optics, 2024, 45(2): 276
    Download Citation