Abstract
1. Introduction
Rotational speed measuring is the most basic and important task in the fields of precision manufacturing, instrumentation, and industrial control[
Commonly used methods to measure rotational speed include the frequency measuring method, periodic measuring method, and frequency–period measuring method, and they calculate the rotational speed by measuring the frequency of the periodic pulse signal caused by the rotor. Because of the pulse quantization error
As a non-contact measurement method, laser Doppler technology can avoid the abovementioned shortcomings, and it has been widely used[
In this paper, a new non-contact method for measuring the instantaneous rotational speed based on the parallel-beam laser Doppler model is proposed. This method can output the rotational speed in real time without measuring the object’s radius. Moreover, the measuring error caused by insufficient installation accuracy (the deviation of the installation angle) of the measuring system can also be avoided. By targeted experiments, the unique characteristics are demonstrated, and we believe that the novel tachometer has great application prospects.
2. Principle of Parallel-Beam Laser Doppler Velocity Measurement System
The optical Doppler effect is the theoretical basis of the laser Doppler velocimeter (LDV), and it indicates the relationship of the measured point’s velocity
In order to solve the problem of measurement error caused by insufficient installation accuracy and radius measurement, a new method called the parallel-beam laser Doppler model for measuring the instantaneous rotational speed is proposed. The structure of the new rotational speed measuring system is shown in Fig. 1.
Figure 1.Structure of the novel rotational speed measuring device.
The light emitted from the laser is collimated by a collimator, and then it is divided by a beam splitter whose reflectivity is 50%. The reflected beam returns in the original direction through the attenuator and the mirror and then gets to the photodetector through the beam splitter, the filter, and the diaphragm, which is the reference light. The transmitted beam is divided into parallel beams by the beam splitter prism with equal optical paths, and they get to the surface of the measured rotating object. The scattered light of the two beams returning in their own original direction is combined as the signal light. Because the angle between one beam and the velocity direction of its corresponding measured point on the rotating object is different from that of another beam, the signal light contains two different frequencies. We can accurately extract two Doppler frequencies by the dual-spectrum-peak-processing circuit used in the measuring system and then calculate the rotational speed. The mixing of the signal light and the reference light on the detector’s photosensitive surface will produce two different Doppler frequencies, and they can be given by
The fast Fourier transform (FFT) is used for signal processing, mainly because this method has a strong ability to extract signals from noise and can accept discontinuous signals[
Figure 2 shows the geometric relationships between the two beams and the measured rotating body.
Figure 2.Geometric relationships of the two beams and the measured rotating body.
According to the geometric relationships contained in Fig. 2, Eqs. (4) and (5) can be further expressed as
So, the difference of two Doppler frequencies generated by the two beams can be given by
From Eq. (8), we can obtain the rotational speed of the measured object:
According to Eq. (9), the rotational speed of the measured object can be calculated only with the distance
3. Experiments
The rotational speed measuring system shown in Fig. 3 is developed according to Fig. 1, and its main performance parameters are shown in Table 1.
Figure 3.Rotational speed measuring system from different perspectives.
Parameter | Value | |
---|---|---|
Laser | Wavelength/nm | 532 |
Power/mW | 50 | |
Mode | Single longitudinal mode | |
Photodetector | APD | |
Sampling frequency/MHz | 100 | |
Measuring range/(deg/s) | 4−8400 | |
Distance | 12 |
Table 1.
It can be seen from Table 1 that the rotational speed measurement range of the measuring system shown in Fig. 3 is 4–8400 deg/s, which is determined by the Doppler frequency range that the tachometer can detect, that is, 20 kHz–50 MHz. Referring to Fig. 2, the diameter of the turntable is 360 mm, and the angle
According to Eq. (10), the range of
In addition, the distance between the parallel beams is 12 mm, and it is measured by the spot position detector[
Sign up for Chinese Optics Letters TOC. Get the latest issue of Chinese Optics Letters delivered right to you!Sign up now
In order to verify the above proposed characteristics of the tachometer designed in this paper, experiments were carried out from four aspects: accuracy evaluation, radius sensitivity analysis, angle sensitivity analysis, and optimization experiment. With the help of the Swiss turntable, an experimental platform was built, as shown in Fig. 3. The rotational speed of the Swiss turntable is very stable, and its accuracy is good to
3.1 Accuracy evaluation experiment
The rotational speed of the turntable is set to 200, 600, and 1200 deg/s, respectively, and, then, the corresponding measured values and the measurement accuracy of the tachometer can be obtained, as shown in Fig. 4.
Figure 4.Results of rotational speed measurement: (a) 200 deg/s, (b) 600 deg/s, and (c) 1200 deg/s.
It can be seen from Fig. 4 that the measurement results of the tachometer fluctuate within a very small range centered on the set value of the turntable. The variances of measurement data are
3.2 Radius sensitivity experiment
The rotational speed of the turntable is set to
Figure 5.Photo of radius sensitivity experiment: (a) a piece of foam, (b) two pieces of foam.
Figure 6.Results of rotational speed measurement: (a) no foam, (b) a piece of foam, and (c) two pieces of foam.
Figures 6(a)–6(c) show the measurement results of the tachometer under three conditions when the turntable’s speed is
The variances of three sets of data are
3.3 Angle sensitivity experiment
The rotational speed of the turntable is also set to
Figure 7.Photo of angle sensitivity experiment.
Figure 8.Results of rotational speed measurement: (a) deflection 0 deg, (b) deflection 5 deg, and (c) deflection 10 deg.
The tachometer is fixed on the chassis through a small tripod, and the initial position of the tachometer’s edge is marked as the baseline position shown in Fig. 7 with a marker pen. Then, the tachometer is rotated slightly in the horizontal plane after loosening the ball joint compression switches on both sides of the tripod, and the edge position can be determined using the marker and protractor until the edge reaches the desired deflection angle position. Next, the ball hinge switch can be closed. In particular, the accuracy requirements for the deflection angle are not very high, because the purpose of the angle sensitivity experiment is to verify that the deflection angle has no effect on the measurement accuracy of the tachometer.
Figures 8(a)–8(c) show the measurement results of the tachometer under three installation angles when the turntable’s speed is set to
3.4 Supplementary optimization of performance evaluation experiments
In order to verify the advantages of the tachometer in terms of accuracy, applicability, and real-time performance more comprehensively, an optimization experiment is carried out: we adjust the rotational speed of the turntable manually as
Sign up for Chinese Optics Letters TOC. Get the latest issue of Chinese Optics Letters delivered right to you!Sign up now
Figure 9.Measurement results under four working conditions.
Figure 9 contains the measurement results under four working conditions, which are the normal measurement state, the sandpaper on the circumference of the turntable, the tachometer deflection
Taking the measurement results at a constant speed of
In order to verify whether the result of the tachometer is accurate compared to the turntable’s setting value, a hand-held tachometer is used. The results are shown in Fig. 10. Because the hand-held tachometer calculates the average rotational speed based on the rotation times (pulse number) of the target paper on the circumference of the turntable in a certain period, it can measure the speed normally during the uniform speed stage, and the measurement results are basically consistent with the novel tachometer. The relative measurement error is the same as the traditional pulse counting method (not high enough). In the acceleration phase, the measurement result of the hand-held tachometer is a series of invalid data points, as shown in Fig. 10 and partially enlarged view No. 1. In addition, the data points marked 2, 3, and 4 are also invalid data points during the acceleration and deceleration phases.
Figure 10.Measurement results of the hand-held tachometer.
Therefore, compared with the traditional rotational speed measurement method, the tachometer in this paper not only improves the measurement accuracy, but also improves the real-time performance of the speed measurement. It is more practical.
Although the accuracy of the novel tachometer is high, and it will not be influenced by the radius fluctuation and the installation angle deviation, other factors that affect the measurement accuracy still exist. To improve the tachometer’s accuracy effectively in the future, we make a simple analysis of these factors as follows.
Considering the influence of air refractive index, Eq. (9) can be transformed into
Taking the full differential of Eq. (11) and performing simple transformation, we can obtain the equation
According to Eq. (12), we can know that the factors that will affect the accuracy of the proposed tachometer include the Doppler frequency spectrum, the change of the refractive index of air, the error of the distance between the parallel beams, and the line width of the laser. The calculation of relative error caused by each factor has been discussed in detail in our previous work[
In addition, the environment vibrations can also affect the accuracy, because they will lead to the relative motion of the measuring system and the measured object. Therefore, we conduct the vibration influence experiment to verify this point, and Fig. 11 is the measurement result under the environmental vibrations (we tap the framework six times randomly and intermittently). The measured rotational speed fluctuates at each beat obviously. Therefore, the necessary vibration suppression measures are needed in future applications.
Figure 11.Measurement result of the vibration influence.
4. Conclusion
A laser Doppler tachometer is developed based on parallel beams in this paper. Compared with the traditional contact method, this tachometer does not need the sensors installed on the measured rotating body directly, which belongs to the non-contact method, and there is no quantization error of the counting pulse in actual usage. So, this tachometer improves the measurement accuracy and convenience. At the same time, the deviation of the measuring system’s installation angle will not have a greater impact on the measurement accuracy when comparing with the common measurement method. In addition, it can output the rotational speed of the rotating body in real time without the angular conversion based on radius, so the measurement error caused by the radius’s fluctuation of the measured rotating body can be avoided. The relative error of the rotational speed measurement is less than 0.06% (
References
[1] H. Wu, C. Chen, C. Li, X. Wang. New method of speed measurement based on PSD. J. Electron. Meas. Instrum., 28, 1033(2014).
[4] Z. Zhou, R. Li, H. Li. Research on high precision measurement system of rotation rate. Meas. Control Technol., 19, 60(2000).
[6] Q. Chen, L. Xue, H. Rao. Rotational speed measurement of ring spinning based on magnetic sensor. Meas. Sci. Technol., 28(2017).
[12] P. Xiao, J. Zhou. Application of referenced light laser Doppler velocimeter in rotation speed measurement of rotator. Chn. Sci. Technol. Inf., 18, 50(2012).
[14] C. Qing, B. Gao, T. Dong, H. Li, S. Yin, C. Jiang, P. Chen. Rotation speed measurement based on self-mixing speckle autocorrelation spectrum. Optik, 208, 164117(2019).
[15] S. He. The micro motor speed measuring system based on laser-Doppler effect(2010).
[16] X. Shen. Laser Doppler Velocimetry and Its Application(2004).
[17] J. Zhou, H. Huang. Application of frequency spectrum refinement and correction technology in laser Doppler velocimeter. Laser Inf., 40, 2(2010).
[19] H. Zhang, Y. Chen, T. Gen, J. Wu, T. Chen. Study on main factors affecting position detection accuracy of four-quadrant detector. Chin. J. Lasers, 42, 1217002(2015).
[20] J. Zhou. Preliminary study on laser Doppler velocimeter for vehicle self-contained navigation(2011).

Set citation alerts for the article
Please enter your email address