Optical coherence tomography (OCT) is an imaging modality proposed in 1990s[1], and it is now broadly applied in various fields, such as ophthalmology[2,3], dermatology[4,5], and gastroenterology[6,7]. One advantage of this technique is its high-resolution imaging, which is achieved with the interference of the broadband spectrum. In Fourier domain OCT systems, the typically used light sources for spectral domain OCT (SDOCT)[8] and swept source OCT (SSOCT)[9] are, respectively, a superluminescent diode (SLD) and a swept source, where an optoelectronic diode is used for the stimulated emission in both situations. Due to the limited bandwidth of the optoelectronic diodes, a typically achieved axial resolution is
For SDOCT, one of the key components for high-quality imaging is the high-resolution spectrometer, in which a collimated beam could be dispersed and focused on a line-array sensor for interference fringes capture. The performance of the spectrometer is determined by the spot size of the different wavelength components on the line-array sensor. The less overlap there is between the wavelength ranges of the two adjacent sensor pixels, the higher fringes contrast. Therefore, many spectrometer designs have recently been proposed to improve the performance. For example, Lee et al. designed and fabricated an F-theta lens and used it in the spectrometer as the Fourier lens (the lens just before the line-array sensor) to correct the field curvature aberration, achieving improved sensitivity fall-off as imaging depth increases[13]. Wu et al. performed a prism and a multi-element scan lens to achieve a high-resolution and wavenumber-linear spectrometer, which could skip fringe resampling during data processing process[14]. Wang et al. used a concave mirror instead of a lens in the spectrometer to improve the transmission efficiency[15], achieving improved signal-to-noise ratio and sensitivity. However, in the above mentioned reports, a complicated optical design, an optical manufacturing, or a system structure is needed, making the system more expensive or optical components alignment more challenging.
In the standard structure of spectrometers, the challenging part is to make the dispersed beam foci after the Fourier lens reach the line-array sensor simultaneously, and one of the commonly used methods is to apply an F-theta lens (which has a flat imaging field) as the Fourier lens, such as Lee et al.’s work[13]. However, an F-theta lens typically costs
In this Letter, we present a low-cost, high-resolution spectrometer design for UHR-OCT, in which multiple standard achromatic lenses with longer focal lengths are combined to suppress SA and to replace the F-theta lens without image-quality degradation, achieving a
2. Method
Figure 1 shows the comparison of simulations with Zemax ray tracing software for the spectrometers with three kinds of Fourier lenses: the single achromatic lens, the F-theta lens, and the combined four achromatic lenses. The effective focal length (EFL) for all three Fourier lenses is 50 mm. In the design, the collimating lens is an achromatic lens with a focal length of 30 mm. The grating has a resolution of 1500 lp/mm with a thickness of 1 mm. All the parameters of the optical lenses are obtained from Thorlabs, US. From the left panel of Fig. 1(a), we can see that the distributions of the foci of the five representative wavelengths (750 nm, 800 nm, 850 nm, 900 nm, and 950 nm) are not flat and could not reach the line-array sensor simultaneously, causing spectral resolution degradation on both sides of the line-array sensor, which agrees with the above-mentioned fundamentals. For the other two settings, the foci of the mentioned five wavelengths reach the line-array sensor almost at the same time, and the performances are comparable to each other. The right panels of Figs. 1(a)–1(c) are the magnified spot diagrams of the five wavelengths on the sensor array plane, and the root-mean-square (RMS) radii of the five diagrams are shown in Table 1. Comparing these values, we can see that the performances of the F-theta lens and the combined four achromatic lenses are close to each other and are much better than the single achromatic lens.
Wavelength (nm) | RMS Radius (µm) | ||
---|---|---|---|
Single Achromatic Lens | F-theta Lens | Combination of Four Achromatic Lenses | |
750 | 156.5 | 15.1 | 26.7 |
800 | 43.4 | 13.7 | 16.6 |
850 | 10.0 | 16.4 | 12.8 |
900 | 36.0 | 20.7 | 13.4 |
950 | 149.6 | 21.4 | 19.5 |
Table 1. The Comparison of the Root-Mean-Square (RMS) Radius at the Five Specific Wavelengths
Figure 1.The simulations with Zemax ray tracing software. (a) The spectrometer with one achromatic lens (EFL = 50 mm) used to replace the F-theta lens. (b) The spectrometer with a typical F-theta lens (EFL = 50 mm, SL50-2P2, Thorlabs, NJ, USA). (c) The spectrometer with the combined 4 achromatic lenses (EFL = 50 mm) used to replace the F-theta lens. The spot diagrams at the five representative wavelengths (0.75 µm, 0.8 µm, 0.85 µm, 0.9 µm, and 0.95 µm) on the line-array sensor plane are shown for comparison in (a)–(c). The RMS radii of the spot diagrams are shown in Table
As illustrated in Fig. 2, we built an SDOCT system in our lab to verify and validate the results of the simulations, in which the light source is an SLD (cBLMD-T-850-HP-I, SUPERLUM, Ireland) centered at 850 nm with a 3 dB bandwidth of 165 nm, achieving an axial resolution of
Figure 2.The schematic of the home-built SDOCT system. PC1, PC2: polarization controllers; DP: dispersion compensation pair; L1–L4: achromatic lenses; CL: combination of four achromatic lenses. The CL is obtained by putting the four lenses together in one single-lens mounting tube.
In this work, the spectrum measured by the home-built spectrometer is calibrated with a method similar to Wang et al.’s work[17]. A piece of cover slip with a thickness of
3. Results and Discussions
Figure 3 shows the results obtained with our home-built SDOCT system. Figure 3(a) shows the normalized power spectral density obtained by our customized high-resolution spectrometer. Figure 3(b) shows the reference plot of the normalized power spectral density from the user manual of the light source. In comparing Figs. 3(a) and 3(b), it can be seen that the spectrum measured by the customized spectrometer generally matches with the reference data, but the intensity of the spectrum on both edges is slightly weaker compared to the reference data. This is caused by the accumulated spectral filtering of the optical components, such as fiber coupler and lenses. Figure 3(c) shows the point spread functions (PSFs) at different depths, and an optical neutral density filter with an OD of 2 was inserted into the sample arm. The different optical path length difference (OPD) was achieved by varying the OPD of the reference arm. The system achieves a sensitivity of
Figure 3.The results of the home-built SDOCT system. (a) The normalized power spectral density obtained by the customized spectrometer. (b) The reference data of the normalized power spectral density from the user manual of light source. (c) The point spread functions (PSFs) at different depths with an optical neutral density filter (OD = 2) inserted in the sample arm. (d) A healthy volunteer’s finger, and the marked position was scanned. (e) A 10 B-scans averaged structural image obtained by the proposed low-cost high-resolution spectrometer. The marked positions are blood vessels.
For spherical lens, SA is unavoidable because off-center light rays are refracted more than those close to the center. To shorten the focal length of spherical lens, the curvatures of the surfaces need to be increased, and the SA is more severe in this case. In this work, four lenses with a focal length of 200 mm are combined to achieve an equivalent lens with an EFL of 50 mm without improving the curvatures of any refractive spherical surfaces, and the obtained performance is comparable to the F-theta lens.
In Fig. 3(c), the full width at half-maximum (FWHM) of the PSFs at different depths slightly varies. This may be caused by the laser diode used in the spectrometer calibration process because the laser diode used in this work is a regular laser diode whose center wavelength drifts as the temperature changes. The calibration accuracy could be improved by using a standard light source instead.
4. Conclusion
In summary, this work presents a low-cost design for the high-resolution spectrometer for ultra-high resolution optical coherence tomography (UHR-OCT), in which four standard achromatic lenses are combined to replace the expensive F-theta lens. Both simulations and in vivo experiments were implemented to test the proposed method, and the obtained results demonstrate that this method could reach a comparable performance to a typical F-theta lens. Because OCT demands relatively expensive clinical devices, we believe this work could help reduce the cost for clinical systems, and it could be helpful to the resources-limited researchers as well.