• Journal of Applied Optics
  • Vol. 45, Issue 5, 896 (2024)
Yuyao LI1,2,*, Chunying PANG1, Yu YANG1, and Xiaoling SUN1
Author Affiliations
  • 1School of Life Science and Technology, Changchun University of Science and Technology, Changchun 130022, China
  • 2Zhongshan Institute of Changchun University of Science and Technology, Zhongshan 528437, China
  • show less
    DOI: 10.5768/JAO202445.0501003 Cite this Article
    Yuyao LI, Chunying PANG, Yu YANG, Xiaoling SUN. Design of optical system for disposable epidural endoscope[J]. Journal of Applied Optics, 2024, 45(5): 896 Copy Citation Text show less
    Structural type of epidural endoscope
    Fig. 1. Structural type of epidural endoscope
    Optical system diagram of disposable epidural endoscope
    Fig. 2. Optical system diagram of disposable epidural endoscope
    MTF curves of disposable epidural endoscope optical system
    Fig. 3. MTF curves of disposable epidural endoscope optical system
    Spot diagram of disposable epidural endoscope optical system
    Fig. 4. Spot diagram of disposable epidural endoscope optical system
    Field curvature and distortion curves of disposable epidural endoscope optical system
    Fig. 5. Field curvature and distortion curves of disposable epidural endoscope optical system
    Pupil ghost image formed by reflection from the 6th to 5th surface
    Fig. 6. Pupil ghost image formed by reflection from the 6th to 5th surface
    Focal ghost image formed by reflection from the 10th to 9th surface
    Fig. 7. Focal ghost image formed by reflection from the 10th to 9th surface
    ParametersSpecification
    Number of pixel328×250
    Pixel size/μm2.5×2.5
    Panel size/mm0.82×0.625
    Wavelength/nmVisible(486.1~656.3)
    F#4.0
    Field of view/(°)90
    System length/mm<1.6
    Distortion/%<20(at 1.0 Field)
    Relative illuminance/%>60(at 1.0 Field)
    Table 1. Design parameters of optical system for epidural endoscope
    CoefficientsS6S7
    a00.860 1590.744 007
    a1−0.136 5580.263 194
    a20.043 941−0.074 506
    a3−0.011 1430.015 127
    a4−2.148 213×10−3−8.066 591×10−4
    a5−7.474 335×10−4−8.922 627×10−4
    Table 2. Aspherical coefficients