- Chinese Optics Letters
- Vol. 22, Issue 4, 041407 (2024)
Abstract
1. Introduction
Due to noncontact, high measurement accuracy, high dynamic measurement range, and the fact that the measurement results can be directly traced back to the length reference, laser interference displacement measurement techniques and instruments have been widely used in the fields of ultraprecision equipment processing and manufacturing. Laser interferometers also play an important role in ocean exploration, geodesy, geophysics, and national security. Laser interferometers in these areas are often faced with harsh external environments, and they are required to have a certain degree of measurement accuracy, which poses a serious challenge to the laser interferometers[1–3].
Taking the laser interferometric absolute gravity meter as an example, to obtain accurate gravity values at different locations in the world, they must perform accurate measurements in field environments to provide data support for scientific research. The current method of ensuring the accuracy of laser interferometers is to keep them warm until they are at a normal temperature, which undoubtedly greatly increases the cost of manpower and time costs[4,5].
As the measurement basis of the laser interferometer, the frequency accuracy, stability, and environmental adaptability of the laser source directly determine its working accuracy under severe environments. Owing to their direct traceability, narrow linewidth, and easily being frequency-stabilized, He–Ne lasers have been widely studied and applied in laser interferometers.
Sign up for Chinese Optics Letters TOC. Get the latest issue of Chinese Optics Letters delivered right to you!Sign up now
At present, He–Ne laser frequency stabilization methods mainly include iodine molecule saturation absorption frequency stabilization and dual longitudinal mode thermal frequency stabilization. Although the iodine saturation absorption frequency stabilization method has high frequency stabilization accuracy, its long preheating time and temperature characteristics of the iodine molecule make it difficult to apply to rapid field measurements. Dual longitudinal mode thermal frequency stabilization has a simple structure and no modulation output, but its working environment is also mainly concentrated in the laboratory or industrial site; otherwise the laser will produce frequency drift or even cannot work properly. Although researchers have proposed techniques such as water cooling and air cooling to ensure a stable thermal environment and reduce the thermal pollution of the laser tube, they have not been able to effectively extend the laser’s operating temperature[6–8].
This study took the dual longitudinal mode thermally stabilized He–Ne laser as the research object, and investigated the drift characteristics of the laser’s frequency with the ambient temperature in the temperature range of
2. Principle
2.1 Preheating setting scheme
The normal working process of the laser can be divided into two parts: preheating and frequency stabilization. The purpose of preheating is to establish approximate thermal equilibrium between the laser and the environment, which is a prerequisite for entering the frequency stabilization control process.
When a laser is free-running in the absence of electric heating film conditions, it reaches thermal equilibrium with the ambient temperature after a period of time and maintains a certain temperature difference that is caused by the self-heating power. As shown in Fig. 1(a), the boundary temperature of the laser tube and the thermal structure is approximately equal (
Figure 1.(a) Thermal equilibrium state; (b) thermal equilibrium temperature of the laser free-running experiment; (c) comparison of the preset temperature and actual temperature of the laser cavity.
Assume that the thermal energy with power
This study used a high- and low-temperature test chamber (temperature adjustment range,
Because the electrothermal film can only increase thermal energy and adjust the cavity length monotonically, to establish a controllable thermal equilibrium state, the laser preheating temperature should be higher than its free-running temperature. Considering the power of the heated film and the limit of the laser tube operating temperature (approximately 70°C), the preheating temperature (
This can effectively solve the problem encountered under the traditional fixed preheating temperature solution, in which it is difficult to establish an effective thermal equilibrium state when the ambient temperature is too low or too high, resulting in the laser not operating properly and accurately.
2.2. Preheating temperature control scheme
To ensure the accuracy of the laser preheating temperature, researchers from the Harbin Institute of Technology have proposed a preheating temperature control scheme based on longitudinal mode-hopping feedback. In this scheme, every time the laser cavity length changes by
After the laser is powered up, the number of longitudinal modes (
Because
In this study, 13 ambient temperature points were selected in the range of
2.3. Nonequilibrium power-locked frequency stabilization scheme
Although the preheating temperature control scheme guarantees frequency repeatability at the same ambient temperature point, the problem of laser frequency drift still exists. This is because the output frequency accuracy depends, to a certain extent, on the laser preheating temperature. The laser output frequency will drift with the difference in the preheating temperature.
The frequency difference of the two modes depends primarily on the mode spacing and is affected by the mode pulling and pushing effects. The laser output beam frequency is
However, the center frequency of the Doppler width varies with the pressure inside the tube with a rate of 18 MHz/Torr[11]. Hence, the Eq. (7) is changed in
Meanwhile, the pressure in the laser tube changes linearly with the temperature, which explains why the laser frequency drifts in the range of
To further verify the output frequency drift with temperature, experiments on beat frequency with an iodine-stabilized laser (ISL) were performed at 13 temperature points, shown in Fig. 1(c). To isolate vibration, the beat frequency device including ISL and the constant temperature box are not in the same room. The data were collected using the frequency counter 53230A (Agilent Technologies, Santa Clara, USA) with a sampling time of 0.1 s. The beat frequency at each temperature point was maintained not less than 3 h; the experimental results are shown in Fig. 2(a). The laser frequency drift peak-to-peak value was approximately 16.2 MHz within
Figure 2.(a) Laser beat frequency at different temperatures; (b) nonequilibrium power-locking principle; (c) variation in laser frequency with error signal.
According to Eq. (6), the frequency difference between neighboring longitudinal modes is
This study used a 150 mm laser tube.
To correct the reference point of laser frequency stabilization more accurately, the relationship between laser frequency and temperature and the relationship between laser frequency and optical power must be determined.
In Fig. 2(a), the laser center frequency drift with temperature followed a relatively clear linear trend. However, to ensure the accuracy of the laser frequency versus temperature model, this study adapted a segmented linear fit to calculate this model, with the temperature interval set as [
Because the commercial He–Ne laser tube parameters cannot be known, a specific expression for the laser gain curve cannot be directly obtained. Therefore, it is necessary to establish a model for the variation in laser frequency with power through experiments:
The laser power
Thus far, this study has established a variation model of laser frequency with ambient temperature and a variation model of laser frequency with error signal.
3. Experiment
3.1. Experiment setup
The structural design scheme is illustrated in Fig. 3(a). The microcontroller unit (MCU) detects the ambient temperature via the external temperature sensor
Figure 3.(a) Structural solution of laser; (b) frequency stabilization system.
Based on the structural scheme, the laser frequency stabilization system was designed and is shown in Fig. 3(b).
In addition, to minimize the error on frequency stabilization control caused by the optical components temperature drift, the optical components used in this study had good temperature stability, for example, true zero-order half-wave plates were used. Moreover, two-element Si PIN photodiodes were used as PDs, which had good performance consistency; its sensitivity at 633 nm did not change with temperature. And the selected circuit components tolerable operating temperature range was concentrated at
3.2. Experiment results
Using the above scheme, six temperature points were selected in the range of
Sampling Time/s | Ambient Temperature/°C | |||||
---|---|---|---|---|---|---|
−20.13 | −9.22 | 8.64 | 16.13 | 28.32 | 40.13 | |
0.1 | 2.44 × 10−10 | 7.51 × 10−11 | 6.39 × 10−11 | 3.29 × 10−10 | 4.40 × 10−10 | 1.17 × 10−10 |
1 | 6.40 × 10−11 | 1.15 × 10−10 | 3.41 × 10−11 | 1.14 × 10−10 | 1.65 × 10−10 | 1.32 × 10−10 |
10 | 3.97 × 10−11 | 3.98 × 10−11 | 5.07 × 10−11 | 3.41 × 10−11 | 5.14 × 10−11 | 4.00 × 10−11 |
100 | 5.28 × 10−11 | 2.53 × 10−11 | 5.22 × 10−11 | 1.58 × 10−11 | 4.37 × 10−11 | 6.74 × 10−11 |
1000 | 6.12 × 10−11 | 2.85 × 10−11 | 3.17 × 10−11 | 2.52 × 10−11 | 5.44 × 10−11 | 2.33 × 10−11 |
Table 1. Allan Variance of Laser Frequency Stabilization at Different Temperatures
Figure 4.Beat frequency with ISL at different ambient temperatures. (a) −20.13°C, (b) −9.22°C, (c) 8.64°C, (d) 16.13°C, (e) 28.32°C, (f) 40.13°C.
Combining Figs. 4(a)–4(f) and Table 1, it can be seen that the laser beat frequency varied by no more than
The frequency plots at the six temperatures are shown in Fig. 5(a) in the form of error bars. The maximum laser frequency deviation
Figure 5.(a) Comparison of nonequilibrium power and equilibrium power frequency stabilization schemes; (b) beat frequencies of the other two lasers at different temperatures.
Based on this method, two other lasers were designed. In Fig. 5(b), the frequency accuracy of these two lasers in the range of
4. Discussion
In the preheating stage, the preheating temperature setting is very important. However, due to the limitation power of electrothermal film, the laser cannot be stabilized at the same temperature within the range of −20°C to 40°C, which is also the reason for the poor frequency accuracy of ordinary lasers. In this study, to ensure that the laser tube had the same thermal balance state at each preheating temperature, we set a fixed temperature difference between the preheating temperature and the ambient temperature so that the PID parameters of the frequency stabilization control program did not need to be modified with the change of the ambient temperature. In addition, there was a nearly linear relationship between the frequency stabilization temperature and the laser frequency in Fig. 2(a). Therefore, we established the relationship model between laser frequency and temperature by curve fitting, but the model accuracy depends on the number of test points and the model building method. The more test points and the more appropriate the modeling method, the more accurate models, but it also brings more work.
For the dual-longitudinal-mode He–Ne lasers, the variation model of laser frequency and optical power has obvious temperature dependence. This is because the Doppler linewidth of the laser gain curve shifts with the temperature of the resonator, resulting in a change in the slope of the gain curve at the equilibrium power point. Most He–Ne laser tubes have this characteristic. Therefore, the proposed nonequilibrium power-locked frequency stabilization scheme has a wide range of applicability in dual-longitudinal-mode thermally stabilized He–Ne lasers. Furthermore, to further improve the frequency compensation accuracy, a more accurate relationship model of laser frequency, light power, and ambient temperature should be established.
5. Conclusion
In conclusion, this study provided a new approach for improving the frequency accuracy of dual-longitudinal-mode thermally stabilized He–Ne lasers. The ambient adaptive preheating temperature setting scheme set the preheating working temperature based on the ambient temperature, which solved the limitations of the traditional fixed preheating temperature scheme and effectively expanded the laser working temperature range. Based on the nonequilibrium power-locked frequency stabilization control scheme, the relationship model between laser frequency and temperature, and the relationship model between laser frequency and optical power, were established through a combination of experiments and theory, which finally improved the frequency accuracy of the laser by 6 times compared with the equilibrium power frequency stabilization scheme.