Abstract
Keywords
Introduction
Biomedical imaging technology has revolutionized life sciences and medical care by providing images with high spatiotemporal resolution, deep penetration, and rich contrast. In recent years, multiscale imaging has been pursued to better understand physiological and pathological information in tissues of interest
Optical ultrasound sensors have compact sizes, broad bandwidth and high sensitivity per unit area, opening up new possibilities for the implementation of multiscale PAI
In the 1970s, a fiber-optic sensor was developed to detect underwater acoustic waves at audible frequencies for defense applications
In recent years, we have developed a fiber laser sensing technology by exploiting the inter-polarization beat note as the sensing signal
Here, we report our progress in the development of flexible fiber-laser-based ultrasound sensing technology and its applications in multiscale photoacoustic imaging. Section Interaction between ultrasound waves and an optical fiber describes the principle of ultrasound interaction with an optical fiber, where both an exact scattering model and an equivalent spring-mass model are illustrated. We particularly emphasize the principle of ultrasound sensitivity tuning by controlling the bending curvature of the fiber sensor. Section Fiber-laser-based ultrasound detection demonstrates the sensor configuration and characterization. Section Photoacoustic microscopy illustrates the optical-resolution, fast-scanning photoacoustic microscopy process using a straight sensor. Section Photoacoustic computed tomography (PACT) shows the circular scanning and linear scanning PACT with a curved fiber-laser sensor with tunable focus. Finally, section Conclusion and outlook provides a conclusion and discusses future developments and applications of fiber-based photoacoustic imaging technology.
Interaction between ultrasound waves and an optical fiber
Scattering model
In acoustics, a bare optical fiber can be treated as a homogeneous elastic cylinder. The transverse profiles of fiber vibrations can be calculated by solving the acoustic wave equations
Exact calculation of the ultrasound response relies on a scattering model. Here, the interaction between the ultrasound wave and the fiber is described as acoustic scattering as a result of the difference in elastic properties. This incident pressure wave excites a vibration mode in the fiber, and the scattered wave varies the acoustic field in the surrounding fluid medium. In this model, the incident wave is decomposed as a linear summation of cylindrical waves with l=0, 1, 2… as input signals. The induced amplitude and phase of the l=0 or 2 modes can be respectively calculated by solving decoupled Helmholtz equations in combination with continuity boundary conditions. Figure 1(a) shows the example of TR21 and TR22 vibration modes when subjected to normally incident planar pressure waves at 22 and 39 MHz, respectively. Despite the profile of the input acoustic waves, only the cylindrical-wave component with l=2 is responsible for the excitation of these vibrations. In ref.
Figure 1.(
An analogous spring-mass oscillator model
The acoustically induced response can be considered as the weak mechanical resonance of the fiber, which is dampened by the acoustic interaction between the fiber and water. This response is analogous to a dampened harmonic oscillator, for example, a spring-mass oscillator, as shown in Fig. 1(b). In this model, the mass m is tied to a spring with a Hook’s coefficient k. The vibration is driven by an external force f(t) and dampened with a coefficient γ. According to Newton’s second law, the frequency response has a Lorentz profile 1/[i(ω0−ω)+ω0γ]. The induced vibration can be characterized by a central frequency at ω0=(k/m)1/2 and a quality factor Q=(m·k)1/2/γ. Now consider the vibration of the optical fiber immersed in water. In the two-dimensional case, for simplicity, the eigenfrequency of a vibrational mode is determined by the elastic and geometric parameters of the fiber, including the density, acoustic velocities, and fiber diameter. Taking an ordinary 125-μm silica optical fiber as an example, the vibrational frequencies of the R01 and TR21 modes are approximately 29 and 22 MHz, respectively. Its vibration is driven by the applied ultrasound and dampened by the surrounding water. The differences between the cylinder vibration and the spring-mass oscillator are as follows: First, the cylinder vibrates in different modes, which can be denoted with an azimuthal order l and radial order n. Among all the vibration modes of an optical fiber, the axially symmetrical modes (with l=0) and torsional-radial modes (with l=2) can induce optical-phase changes. Second, the frequency response of the cylinder vibration is expressed by Bessel functions of kart, where ka and rt denote the acoustic wavenumber and radial position. Therefore, this response somewhat deviates from the well-known Lorentz curve.
Response of a straight fiber to a spherical pressure wave
It is worth noting that the spherical ultrasound waves generated by the individual absorbers are detected with PAI, rather than the plane waves. Here, the acoustic response of an ideal line detector to a spherical wave is discussed. As shown in Fig. 1(c), a point source S generates an acoustic wave with a frequency fa. Suppose the acoustic wave is totally absorbed by the line detector placed at a distance d. The acoustic pressure propagation can be written as p(ω, r)=exp(−ikar)/r, where r=(d2+z2)1/2 is the propagation distance, and z represents the longitudinal position. The spherical wavefront arrives at the line detector with different phases. The response of the line detector can be expressed as an integration: R =
Response of a curved fiber to a spherical pressure wave
Compared with rigid piezoelectric transducers, optical fibers can be curved into different shapes thanks to the great flexibility of the optical fiber in geometry. The above analysis suggests that the variation in the acoustic phase along with the fiber largely determines the sensitivity and the frequency response. This implies a potential degree-of-freedom in the sensitivity enhancement by bending the fiber to coincide with the spherical wavefront
Figure 2.(
where Г is a coefficient accounting for the absorption or scattering experienced by the sound when traveling from the source S to the fiber and A(ω) is the normalized frequency response of a straight fiber to plane ultrasound waves. The fiber length L, defined as the ultrasound-sensitive zone, is approximately equal to the cavity length of the laser sensor. By varying the locations (x, y, z) of the point source S, the three-dimensional spatial response to ultrasound can be plotted, as shown in Fig. 2(b, c, e, f). Figure 2(b, c) suggests the decaying sensitivity of the straight sensor with increased source-detector distance, described as the phase cancellation effect in the next Section Fiber-laser-based ultrasound detection. By contrast, the curved sensor has a significantly different sensitivity profile, as shown in Fig. 2(e, f). It has maximal sensitivities to the acoustic sources around the focus spot (at x=30 mm) with a 5-fold enhancement, owing to the coincidence between the spherical wavefront and the curved sensor. Notably, the fiber sensor also has a large view angle, yielding an extended field of view over the x-z plane. Therefore, the focal length and the sensitivity profile can be continuously adjusted by bending the fiber sensor into different curvatures, which opens up the possibility of multi-depth imaging. Additionally, the focused acoustic beam has a width of approximately 500 μm in the z direction endowing the sensor with sectioning capability for PAI.
Fiber-laser-based ultrasound detection
The ultrasound sensing system
In recent years, ultrasound sensors based on optical resonance have been developed, including planar and fiber-based Fabry-Perot sensors, in-fiber phase-shifted gratings, and polymer microring resonators
Notably, the dual-polarization operation is important for ultrasound detection. For the l=2 vibration modes, the x- and y-polarized modes present equal but opposite refractive-index changes. This birefringence change can be effectively measured by the laser sensor. By contrast, these two modes have almost the same response to thermal, vibrational and other low-frequency perturbations. Therefore, common-mode noise cancelation can ensure a high SNR for ultrasound detection.
Figure 3(a) schematically shows ultrasound detection with a fiber-laser-based sensor
Figure 3.(
where c denotes the light speed in vacuum, n0 represents the refractive index of fiber silica, and λ is the lasing wavelength. The pressure wave can deform the fiber, induce a birefringence change dB and modulate the beat frequency, as shown in Fig. 3(b) and expressed as
The acoustically modulated beat signal can be measured via I/Q demodulation. Briefly, the beat signal is mixed with two quadrature radio frequency signals, which have the same frequency and 90-degree phase offset. After mixing, the two baseband signals I and Q are used to recover the modulated signal. The phase information can be extracted via ϕ = arctan(Q/I). The variation in the frequency of the beat signal is then recovered by taking the derivative of the phase.
Sensor characterization
Figure 4.(
Bandwidth
Figure 4(a, b) exhibits the temporal waveform and the corresponding spectrum of the same photoacoustic signal. The central frequency of this mechanic mode is ~22 MHz, and the –6 dB bandwidth is ~70%. As described in the previous Section Interaction between ultrasound waves and an optical fiber, the central frequency depends on the elastic properties of the fiberglass and the diameter of the fiber (~125 µm). The bandwidth is determined by the difference in acoustic properties between the fiber and the surrounding water. The working bandwidth can be extended by etching the fiber with hydrofluoric acid. For example, a 65-µm sensor provides a two-fold bandwidth extension, and the central frequency shifts to approximately 40 MHz.
Noise-equivalent pressure
Figure 4(c) shows the measured SNR at different fiber-laser powers with a fixed peak-to-peak acoustic pressure of the incident planar ultrasound wave at 75.2 kPa. The SNR is mainly limited by the noise figure of the photodetector. In the high-input-power regime, the dominant noise source is the shot noise, and the SNR is approximately proportional to the square root of the optical power. Therefore, the incident optical power is amplified with an erbium-doped fiber amplifier to have the photodetector work in the shot noise limited region. Eventually, saturation of the photodetector will limit optical power. When the input power on the photodiode is 30 mW, the NEP is ~43.6 Pa with a 50-MHz acquisition bandwidth
Stability
Figure 4(d) shows the stability test result of the sensor by measuring the photoacoustic signals from two discrete absorbers excited by using a linearly scanning pulsed laser. The 30-min measurement results show no noticeable variation in the sensitivity. We also observed that even with vibration or temperature changes, the sensor remains stable with high sensitivity. We even scanned the fiber sensor back and forth with a maximum velocity of 10 mm/s with respect to an ultrasound source and found that the sensor can still maintain a stable output. The good stability is also a result of the common noise cancelation, taking advantage of the beat-note based demodulation. The acoustically induced lasing frequency change can also be demodulated based on imbalanced interferometry, but a servo loop for frequency locking is needed.
The performances of the present sensor are compared with the piezoelectric and optical ultrasound sensors in Table 1. This sensor has comparable sensitivities to the existing optical sensors. In addition, the sensor has presented a great flexibility in sensor geometry, making it suitable for both photoacoustic microscopy and computed tomography. Notably, new microlasers have been recently developed with biocompatible materials and can even be implemented into human cells
Sensor type | Material | Bandwidth
| NEP
| Demodulation
| Common
| |
N.R.: not reported. | ||||||
Piezoelectric
| Spherically focused
| Piezoceramic | 50 MHz | 0.2 | Voltage | PAM |
Needle (Precision acoustics) | PVDF | N.R. | 14.4 | Voltage | N.R. | |
Optical ultrasound
| Micro-ring | Polystyrene | 350 | 5.61 | CW interferometry | PAM |
Fabry-Pérot | PET/Parylene C SU8 | 20 | 44.7 | CW interferometry | PACT | |
Epoxy | 2.8 | 1.6 | CW interferometry | PAM | ||
π-FBG | Silica | 16 | 25 | CW interferometry | PAM | |
Table 1. Comparison of piezoelectric and optical ultrasound transducers.
Photoacoustic microscopy (PAM)
Photoacoustic microscopy (PAM) focuses a pulsed laser beam on biological tissue and detects ultrasound waves by using a coaligned focused piezoelectric transducer. The scanning head raster scans over the plane orthogonal to the laser incidence direction to form a two-dimensional image. Additionally, the depths of the absorbers can be retrieved at acoustic resolution based on time-to-space mapping, forming a three-dimensional image. Despite its high spatial resolution and structural/functional imaging capability, PAM is limited by the mechanical scanning speed of the apparatus. A fast-scanning PAM apparatus can be built by using a straight fiber-laser-based ultrasound sensor, taking advantage of its high sensitivity and large field-of-view (FOV). This section reviews fiber-based PAM for in vivo structural and functional imaging at a cellular resolution. Figure 5(a) shows the schematic of the PAM apparatus. Here, the fiber sensor remains stationary, while the optical beam is scanned by using a galvanometric scanner in air to achieve a high speed. A pulsed laser beam is focused on the absorptive sample with an optical-diffraction-limited spot size. The FOV is first characterized before in vivo imaging. Figure 5(b) shows the measured field of view at d=1 mm between the laser fiber sensor and a point-like photoacoustic source. The source is raster scanned with respect to the laser fiber sensor in the x-y plane. The sensitivity profile along the fiber direction is ~2 mm, determined by the longitudinal intensity profile of the intracavity light. The sensitivity profile normal to the fiber direction, in contrast, depends on the radial orientation of the fiber laser sensor in proportion to cos(2θ), where θ represents the orientation angle with respect to one of the fiber principal axes. The ultrasonic response reaches its maximum in the fiber principle axis directions. The full angle at half maximum is 60 degrees. As a result, the effective field of view normal to the fiber direction scales with distance d. The field of view (FOV) presents an effective area (at half maximum) of 1.57 mm2 at d=1 mm with a nearly elliptical profile.
Figure 5.(
Figure 5(c) shows the result of hemodynamic imaging, with a scanning rate increased to 2 Hz and a FOV of 2×2 mm2
Imaging of oxygen saturation (sO2) can provide invaluable information on the functionality and metabolism of living tissues. An sO2 map describes the processes of oxygen transport and consumption in biological tissues, based on which the functionality of the cardiovascular system can be evaluated. Brain functional imaging can be performed with the fiber-laser-based PAM apparatus paired with a multiwavelength laser source. A dual- or multiwavelength source can be built based on stimulated Raman scattering (SBS). Briefly, a nanosecond 532-nm laser is split into two beams, one of which is injected into a high-nonlinearity fiber to generate a 558-nm component via SBS (as the second-order Stokes wave). The other beam transmits through a large-mode-area fiber for time delay without wavelength transition. These two beams are then coupled back into a common fiber and focused on the biological tissue to generate pulse trains with a controllable time interval. The optical absorption of blood at the visible band comes from the mixture of HbR and HbO2, whose concentrations are denoted as CHbR and CHbO2, respectively. The oxygen saturation is defined as the ratio between them, expressed as sO2=CHbO2/(CHbO2+CHbR). Assuming that the amplitude of the photoacoustic signal is proportional to the absorption coefficient, we can write
where PA532 and PA558 represent the photoacoustic amplitudes excited by the dual-color laser pulses. The elements in the 2-by-2 matrix denote the molar absorption coefficients, showing the difference in the absorption spectrum between HbR and HbO2. The sO2 value at each pixel can be derived with the measured photoacoustic amplitudes and the given coefficient matrix. Figure 5(e) shows the cortex blood vessels obtained with a single wavelength laser. In contrast, an sO2 image can be obtained by performing spectral unmixing based on Eq. (4) at each position, as shown in Fig. 5(f). The oxygen saturation levels are represented in different colors, and we are capable of distinguishing the pile surface arteries (red) and veins (yellow).
By controlling the time delay of the ultrasonically modulated laser light, a photoacoustic signal can be measured multiple times to enhance the signal-to-noise ratio
Figure 6.(
Photoacoustic computed tomography (PACT)
The straight sensor showed great potential for in vivo OR-PAM imaging of vasculature in mouse ears. As the sensor sensitivity decayed with increasing working distance, the mouse ear needed to be placed close to the fiber surface during the imaging process. This phenomenon results from the wavelength-scale equivalent interaction length of a straight fiber for spherical ultrasound waves and limits the potential for deep tissue imaging. This issue was addressed by bending the fiber to realize lens-free focused ultrasound sensor with greatly increased sensitivity at long working distance. The focus capability was obtained by simply bending the straight FUS along its axial direction to conform the spherical wavefront of ultrasound. At a curvature radius of 30 mm, the focused FUS showed a ~5 times higher sensitivity compared with the straight one, and was demonstrated for in vivo PACT imaging of a zebrafish and a mouse brain, as shown in Fig. 7.
Figure 7.(
The imaging was performed in a rotary-scanning manner, as shown in Fig. 7(a)
By designing a focus-tuning holder, we further exploits the flexibility of optical fiber to dynamically tune the focus of the sensor by bending the fiber into different curvature radii for multi-depth imaging. The small diameter of the fiber laser also provides a wide acceptance angle and the focusing capability in elevational direction for high-resolution cross section imaging, making the focused fiber laser an ideal candidate for developing a high-resolution linear-scanning PACT system, which allows easy access to patient body trunks and high compatibility with handheld ultrasound probes for preclinical and clinical applications. Figure 8(a) shows the photograph of the customized holder, which can continuously tune the focal depth of the sensor from 22 to 50 mm by simply moving the slide block. Bending the fiber can induce a shift in the carrier frequency of the fiber sensor, but the signal-to-noise ratio is hardly changed. To realize a 2D PACT, the curved fiber laser is kept in the y-z plane, and the holder is scanned by using a motor-driven translation stage in the x direction, as shown in Fig. 8(b). The biological sample is illuminated by using a 532-nm pulsed laser with an expanded beam aperture. The optically induced ultrasound waves are detected at each scanning position and then reconstructed based on the back-projection algorithm. The multi-depth imaging capability was verified by imaging a group of human hairs parallelly aligned with a spacing of 3 mm along the z axis in a block of agar, as shown in Fig. 8(c). The capability of high-resolution small-animal PACT was studied by imaging the abdominal subcutaneous vasculature of a rat (Sprague-Dawley; weight: 150 g) anesthetized with 1.5% isoflurane. The radiant exposure was 3 mJ/cm2, much lower than the ANSI limit. Figure 8(d) shows the maximum amplitude projection image of rat vasculature in a 12 mm×20 mm area, in accordance with the photograph in Fig. 8(e). The in-plane spatial resolution was characterized as 150 μm×80 μm, determined by the bandwidth and the angle-of-view of the fiber laser sensor. The high resolution of the system allows clear observation of vessels with diameters varying from 90 to 400 μm. It is worth to note that the current radiant exposure is ~ 7 times lower than the ANSI limit. A high-power pulsed laser at near infrared wavelength could be employed for high-contrast, high-penetration 3D in vivo imaging in the future.
Figure 8.(
Conclusion and outlook
In conclusion, we have developed an ultrasound sensing technology based on a flexible FUS. The ultrasound pressure is detected by reading out the beat frequency at radio frequencies. The sensor presents outstanding performance in sensitivity, bandwidth and stability. More importantly, the spatial ultrasound response can be freely tuned by simply bending the fiber enabling multiscale PAI with a single fiber laser sensor. Based on this unique flexibility of the sensor, multiscale photoacoustic imaging including micro-scale resolution PAM and deep-tissue PACT have been demonstrated for in vivo small animals imaging. The in vivo imaging results suggest that the fiber-laser based photoacoustic imaging is promising to the development of various modalities including PA microscopy, endoscopy and tomography for both life science research and medical diagnostics. The potential future developments and applications of fiber-based photoacoustic imaging technology are provided as following:
Firstly, the fiber laser sensor may enable miniature head-mounted microscopy for brain activity monitoring. Two- or multiphoton microscopes have been developed with an all-fiber configuration. These microscopes can resolve single neurons in response to a stimulus. However, the field of view is rather limited. Small PAM apparatus has been created with a small piezoelectric transducer, but the weight is typically above 20 grams, which is hardly bearable for small animals. Fiber-laser-based PAM can offer a possible strategy of using a lightweight head-mounted microscope to monitor brain activities based on neurovascular couplings.
Secondly, an all-fiber photoacoustic endoscopy can be achieved with the sensor, equipped with another fiber for the guidance of excitation light pulses. The development of photoacoustic endoscopy is currently limited by the tradeoff between device size and imaging performance, such as sensitivity and spatial resolution. The fiber laser sensor inherently offers a side-view detection with high sensitivity, making it suitable for the miniaturization of endoscopy.
Finally, flexible PACT can be integrated to hand-held scanners for the cancer screening in preclinical/clinical applications. To increase the imaging speed of our previously reported FUS-based PACT, a fiber laser sensor array with multiple parallel elements can be built considering the intrinsic wavelength multiplexing capability of optical fibers. Specifically for the FUS, frequency and time multiplexing techniques are preferred to form a large-scale sensor array with a single RF demodulation unit. Such a system would not only speed up the imaging process but also significantly reduce the cost for preclinical/clinical applications.
References
[1] T Niendorf, L Frydman, M Neeman, E Seeliger. Google maps for tissues: multiscale imaging of biological systems and disease. Acta Physiol, 228, e13392(2020).
[2] LV Wang. Multiscale photoacoustic microscopy and computed tomography. Nat Photonics, 3, 503-509(2009).
[3] YJ Yao, AA Kaberniuk, L Li, DM Shcherbakova, RY Zhang, et al. Multiscale photoacoustic tomography using reversibly switchable bacterial phytochrome as a near-infrared photochromic probe. Nat Methods, 13, 67-73(2016).
[4] A Taruttis, V Ntziachristos. Advances in real-time multispectral optoacoustic imaging and its applications. Nat Photonics, 9, 219-227(2015).
[5] JJ Yao, LV Wang. Photoacoustic microscopy. Laser Photonics Rev, 7, 758-778(2013).
[6] L Li, LR Zhu, C Ma, L Lin, JJ Yao, et al. Single-impulse panoramic photoacoustic computed tomography of small-animal whole-body dynamics at high spatiotemporal resolution. Nat Biomed Eng, 1, 0071(2017).
[7] SM Maswadi, BL Ibey, CC Roth, DA Tsyboulski, HT Beier, et al. All-optical optoacoustic microscopy based on probe beam deflection technique. Photoacoustics, 4, 91-101(2016).
[8] E Leiss-Holzinger, J Bauer-Marschallinger, A Hochreiner, P Hollinger, T Berer. Dual modality noncontact photoacoustic and spectral domain OCT imaging. Ultrason Imaging, 38, 19-31(2016).
[9] P Hajireza, W Shi, K Bell, RJ Paproski, RJ Zemp. Non-interferometric photoacoustic remote sensing microscopy. Light Sci Appl, 6, e16278(2017).
[10] HH Li, F Cao, YY Zhou, ZP Yu, PX Lai. Interferometry-free noncontact photoacoustic detection method based on speckle correlation change. Opt Lett, 44, 5481-5484(2019).
[11] TX Wang, R Cao, B Ning, AJ Dixon, JA Hossack, et al. All-optical photoacoustic microscopy based on plasmonic detection of broadband ultrasound. Appl Phys Lett, 107, 153702(2015).
[12] XJ Zhu, ZY Huang, GH Wang, WZ Li, D Zou, et al. Ultrasonic detection based on polarization-dependent optical reflection. Opt Lett, 42, 439-441(2017).
[13] F Yang, W Song, CL Zhang, CJ Min, H Fang, et al. Broadband graphene-based photoacoustic microscopy with high sensitivity. Nanoscale, 10, 8606-8614(2018).
[14] JA Guggenheim, J Li, TJ Allen, RJ Colchester, S Noimark, et al. Ultrasensitive Plano-concave optical microresonators for ultrasound sensing. Nat Photonics, 11, 714-719(2017).
[15] A Rosenthal, S Kellnberger, D Bozhko, A Chekkoury, M Omar, et al. Sensitive interferometric detection of ultrasound for minimally invasive clinical imaging applications. Laser Photonics Rev, 8, 450-457(2014).
[16] H Li, BQ Dong, Z Zhang, HF Zhang, C Sun. A transparent broadband ultrasonic detector based on an optical micro-ring resonator for photoacoustic microscopy. Sci Rep, 4, 4496(2014).
[17] C Zhang, SL Chen, T Ling, LJ Guo. Imprinted polymer microrings as high-performance ultrasound detectors in photoacoustic imaging. J Lightwave Technol, 33, 4318-4328(2015).
[18] AP Jathoul, J Laufer, O Ogunlade, B Treeby, B Cox, et al. Deep in vivo photoacoustic imaging of mammalian tissues using a tyrosinase-based genetic reporter. Nat Photonics, 9, 239-246(2015).
[19] R Shnaiderman, G Wissmeyer, O Ülgen, Q Mustafa, A Chmyrov, et al. A submicrometre silicon-on-insulator resonator for ultrasound detection. Nature, 585, 372-378(2020).
[20] P Nash. Review of interferometric optical fibre hydrophone technology. IEE Proc - Radar, Sonar Navigat, 143, 204-209(1996).
[21] GA Cranch, PJ Nash, CK Kirkendall. Large-scale remotely interrogated arrays of fiber-optic interferometric sensors for underwater acoustic applications. IEEE Sens J, 3, 19-30(2003).
[22] R DePaula, L Flax, J Cole, J Bucaro. Single-mode fiber ultrasonic sensor. IEEE J Quantum Electron, 18, 680-683(1982).
[23] L Flax, JH Cole, Paula De, JA Bucaro. Acoustically induced birefringence in optical fibers. J Opt Soc Am, 72, 1159-1162(1982).
[24] S Foster, A Tikhomirov, M Milnes, Velzen van, G Hardy. A fiber laser hydrophone. Proc SPIE, 5855, 627-630(2005).
[25] BO Guan, L Jin, Y Zhang, HY Tam. Polarimetric heterodyning fiber grating laser sensors. J Lightwave Technol, 30, 1097-1112(2012).
[26] BO Guan, L Jin, LH Cheng, YZ Liang. Acoustic and ultrasonic detection with radio-frequency encoded fiber laser sensors. IEEE J Sel Top Quantum Electron, 23, 5601712(2017).
[27] T Berer, IA Veres, H Grün, J Bauer-Marschallinger, K Felbermayer, et al. Characterization of broadband fiber optic line detectors for photoacoustic tomography. J Biophotonics, 5, 518-528(2012).
[28] X Bai, YZ Liang, HJ Sun, L Jin, J Ma, et al. Sensitivity characteristics of broadband fiber-laser-based ultrasound sensors for photoacoustic microscopy. Opt Express, 25, 17616-17626(2017).
[29] X Bai, YM Qi, YZ Liang, J Ma, L Jin, et al. Photoacoustic computed tomography with lens-free focused fiber-laser ultrasound sensor. Biomed Opt Express, 10, 2504-2512(2019).
[30] YZ Liang, L Jin, LD Wang, X Bai, LH Cheng, et al. Fiber-laser-based ultrasound sensor for photoacoustic imaging. Sci Rep, 7, 40849(2017).
[31] YZ Liang, JW Liu, L Jin, BO Guan, LD Wang. Fast-scanning photoacoustic microscopy with a side-looking fiber optic ultrasound sensor. Biomed Opt Express, 9, 5809-5816(2018).
[32] M Schubert, L Woolfson, IRM Barnard, AM Dorward, B Casement, et al. Monitoring contractility in cardiac tissue with cellular resolution using biointegrated microlasers. Nat Photonics, 14, 452-458(2020).
[33] YZ Liang, L Jin, BO Guan, LD Wang. 2 MHz multi-wavelength pulsed laser for functional photoacoustic microscopy. Opt Lett, 42, 1452-1455(2017).
[34] YY Zhou, SY Liang, MS Li, CB Liu, PX Lai, et al. Optical-resolution photoacoustic microscopy with ultrafast dual-wavelength excitation. J Biophotonics, 13, e201960229(2020).
[35] YZ Liang, JW Liu, LD Wang, L Jin, BO Guan. Noise-reduced optical ultrasound sensor via signal duplication for photoacoustic microscopy. Opt Lett, 44, 2665-2668(2019).
[36] X Bai, J Ma, X Li, L Jin, BO Guan. Focus-tunable fiber-laser ultrasound sensor for high-resolution linear-scanning photoacoustic computed tomography. Appl Phys Lett, 116, 153701(2020).

Set citation alerts for the article
Please enter your email address