• Optics and Precision Engineering
  • Vol. 30, Issue 13, 1542 (2022)
Zhen CHANG1,2, Yu WANG2,*, Fang LIN1, Xin ZHAO1..., Shujua HUANG1 and Fuqi SI1|Show fewer author(s)
Author Affiliations
  • 1Key Laboratory of Environmental Optical and Technology, Anhui Institute of Optics and Fine Mechanics, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei23003, China
  • 2Information Materials and Intelligent Sensing Laboratory of Anhui Province, Institutes of Physical Science and Information Technology, Anhui University, Hefei30039, China
  • show less
    DOI: 10.37188/OPE.20223013.1542 Cite this Article
    Zhen CHANG, Yu WANG, Fang LIN, Xin ZHAO, Shujua HUANG, Fuqi SI. Measurement of pixel full-well parameters of detector used on spaceborne spectrometer[J]. Optics and Precision Engineering, 2022, 30(13): 1542 Copy Citation Text show less
    Schematic of LED-point test method
    Fig. 1. Schematic of LED-point test method
    Schematics of detector imaging system test of spaceborne detector
    Fig. 2. Schematics of detector imaging system test of spaceborne detector
    Test results of standard deviation-mean logarithm PTC in full frame mode
    Fig. 3. Test results of standard deviation-mean logarithm PTC in full frame mode
    Test results of variance-mean PTC in full frame mode
    Fig. 4. Test results of variance-mean PTC in full frame mode
    Test images of LED-point method
    Fig. 5. Test images of LED-point method
    Test results of LED-point method
    Fig. 6. Test results of LED-point method
    Test results of 4 different pixels. (a1),(b1),(c1),(d1) Trend map of the 9 LED pixel values under different light intensities; (a2),(b2),(c2),(d2) The inflexion point of the curve obtained by nonlinear fitting the overflow pixel DN value data, and the corresponding DN value of the point in the center pixel curve
    Fig. 7. Test results of 4 different pixels. (a1),(b1),(c1),(d1) Trend map of the 9 LED pixel values under different light intensities; (a2),(b2),(c2),(d2) The inflexion point of the curve obtained by nonlinear fitting the overflow pixel DN value data, and the corresponding DN value of the point in the center pixel curve
    Test result of LED point using traditional FWC test method
    Fig. 8. Test result of LED point using traditional FWC test method
    ParametersValue
    Readout Noise(DN)4.762
    System Gain(DN/e-0.011 05
    FWC(ke-817.7
    Table 1. PTC test results of detector in satellite payload
    Pixel no.Saturated values/DNMean value/DNDeviation/%FWC/ke-
    19 5629 550.4(减去偏置后对应满阱817.013 ke-0.121 5818.063
    29 5870.383 2820.326
    39 6070.592 6822.136
    49 5010.517812.543
    59 4950.58812
    Table 2. Test results of detector with LED-point test method.
    Zhen CHANG, Yu WANG, Fang LIN, Xin ZHAO, Shujua HUANG, Fuqi SI. Measurement of pixel full-well parameters of detector used on spaceborne spectrometer[J]. Optics and Precision Engineering, 2022, 30(13): 1542
    Download Citation