• Infrared and Laser Engineering
  • Vol. 51, Issue 7, 20220431 (2022)
Liu Zhang, Bonan Li, Yongnan Lu*, Yangyang Zou, and Tailei Wang
Author Affiliations
  • College of Instrument Science and Electrical Engineering, Jilin University, Changchun 130021, China
  • show less
    DOI: 10.3788/IRLA20220431 Cite this Article
    Liu Zhang, Bonan Li, Yongnan Lu, Yangyang Zou, Tailei Wang. Optical system design of spaceborne CO2 imaging spectrometer based on Offner convex grating[J]. Infrared and Laser Engineering, 2022, 51(7): 20220431 Copy Citation Text show less
    Schematic diagram of Offner grating spectrometer
    Fig. 1. Schematic diagram of Offner grating spectrometer
    Meridian imaging
    Fig. 2. Meridian imaging
    Sagittal imaging
    Fig. 3. Sagittal imaging
    Schematic of the design process
    Fig. 4. Schematic of the design process
    Initial structure diagram
    Fig. 5. Initial structure diagram
    Spot diagram of the initial structure with the wavelength of (a) 1.594 μm, (b) 1.6065 μm, (c) 1.619 μm; MTF of the initial structure with the wavelength of (d) 1.594 μm, (e) 1.6065 μm, (f) 1.619 μm
    Fig. 6. Spot diagram of the initial structure with the wavelength of (a) 1.594 μm, (b) 1.6065 μm, (c) 1.619 μm; MTF of the initial structure with the wavelength of (d) 1.594 μm, (e) 1.6065 μm, (f) 1.619 μm
    (a) Structure of optimized optical system; (b) Structure of optical-mechanical
    Fig. 7. (a) Structure of optimized optical system; (b) Structure of optical-mechanical
    Spot diagram of the optimized system with the wavelength of (a) 1.594 μm, (b) 1.606 5 μm, (c) 1.619 μm; MTF of the optimized system with the wavelength of (d) 1.594 μm, (e) 1.606 5 μm, (f) 1.619 μm
    Fig. 8. Spot diagram of the optimized system with the wavelength of (a) 1.594 μm, (b) 1.606 5 μm, (c) 1.619 μm; MTF of the optimized system with the wavelength of (d) 1.594 μm, (e) 1.606 5 μm, (f) 1.619 μm
    Geometric encircled energy
    Fig. 9. Geometric encircled energy
    Optimize keystone image
    Fig. 10. Optimize keystone image
    Optimize smile image
    Fig. 11. Optimize smile image
    ParameterValue
    Object space NA0.2
    Spectral range/nm1594-1619
    Spectral resolution/nm0.1
    Slit length/mm30
    Pixel dimension15 μm×15 μm
    MTF≥0.7
    Keystone and smile/pixel<10%
    Table 1. Principal design indicators of the system
    SurfaceRadius/mmThickness/mmMaterialLines/mm
    Object-415.43--
    1−422203.59Mirror-
    Stop−214.58203.59Mirror550
    3−415.5203.59Mirror-
    ImageInfinity219.03--
    Table 2. Structure parameter of initial optical
    SurfaceRadius/mmThickness/mmDiameter/mmGlassLines/mm
    Object-650.655Φ200 --
    1300.5760Φ375 H-QF3-
    2373.35999.812---
    3−1703.59−978.79Φ500 Mirror-
    Stop−1004.4−978.79Φ120 Mirror550
    5−1920.66−999.812Φ660 Mirror-
    6373.35−60Φ600 H-FK61 B-
    7300.57−754---
    ImageInfinity----
    Table 3. Structure parameter of optimization system
    Liu Zhang, Bonan Li, Yongnan Lu, Yangyang Zou, Tailei Wang. Optical system design of spaceborne CO2 imaging spectrometer based on Offner convex grating[J]. Infrared and Laser Engineering, 2022, 51(7): 20220431
    Download Citation